FujiFilm PB-30 Operation Manual page 18

Balloon controller
Table of Contents

Advertisement

Chapter 3 Name and Function of Each Part
Toggle switch 2 (Black: Endoscope side)
2
Pause switch 2 (Endoscope side)
3
Stop switch
4
Pause switch 1 (Over-tube side)
5
RC connector
6
18
This switch toggles between air supply and air discharge to/from the balloon on the
endoscope side.
When the power is turned on, the inside light turns on and air is discharged from the
balloon.
When this switch is pressed while the inside light is on, the outside light turns on and air is
supplied to the balloon. When this switch is pressed while the outside light is on, the inside
light turns on and air is discharged from the balloon.
[Note] This switch is disabled if the balloon switch 2 on the main unit is off.
If you press this switch during air supply or discharge to/from the balloon on the endoscope
side, air supply or discharge is paused. The switch lights up during pausing.
[Note] When immediately after the power source ON, air is not supplied or discharged until
the line pressure reaches -6.0 kPa.
While the buzzer is sounding, if you press this switch, the buzzer stops sounding.
If you press this switch during air supply or discharge to/from the balloon on the over-tube
side, air supply or discharge is paused. The switch lights up during pausing.
[Note] When immediately after the power source ON, air is not supplied or discharged until
the line pressure reaches -6.0 kPa.
Connects to the remote terminal of the main unit.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents