Conditions Of Disinfection - FujiFilm PB-30 Operation Manual

Balloon controller
Table of Contents

Advertisement

Chapter 9 Chemical Disinfection

9.2 Conditions of Disinfection

56
Disinfect the tube and balloon according to the specific methods
described in this manual. There is a risk of damage to the equipment.
The tube and balloon have not been disinfected and sterilized.
When using them for the first time, use the level of disinfection or sterilization suitable to
the application.
Legally marketed 2.4 to 3.5% glutaraldehyde solutions can be used with the tube and
balloon for disinfection when used in accordance with the manufacturer's instructions.
Contact your local FUJIFILM dealer for names of compatible glutaraldehyde solutions.
Refer to the chemical solution manufacturer's instructions regarding preparation, use and
exposure conditions.
Check the expiration date of chemical solution before use. Do not use a chemical solution
beyond its expiration date.
Do not use a chemical solution whose efficacy is not sufficient. Check the chemical
solution's efficacy before use according to the manufacturer's recommendation (such as
using a test strip), referring to the instructions of chemical solution.
Ensure that the tube and balloon are completely immersed during disinfection. If the tube
and balloon are not completely immersed, disinfection may not be achieved.
The balloon has not been pre-sterilized. When using it for the first time, use the level of
reprocessing suitable to the application, in accordance with Chapter 8 "Cleaning," Chapter
9 "Chemical Disinfection," and Chapter 10 "Sterilization."
The balloon is intended for single patient use. Do not reuse it in order to prevent infection.
CAUTION