Chapter 12 Troubleshooting

12.3 Error Codes

Error code
72
If any abnormality occurs, the measured pressure and the error code are displayed
alternately on the pressure display.
Indication method
Alternate indication
of the error and the
pressure for the
balloon in question
Alternate indication
at pressure
indicators 1 and 2
Alternate indication
of the error and the
pressure for the
balloon in question
Possible cause
Tube disconnection
detected at power-on
Abnormal balloon
pressure detected
Tube disconnection
detected during air
supply
Broken balloon detected
Tube disconnection
detected during air
discharge
Abnormality with the
fan in the main unit
Abnormality in the
pressure sensor inside
the main unit
Action to take
Clear the buzzer using the STOP switch.
Turn the power off, connect the tube and
then turn the power back to on.
Clear the buzzer using the STOP switch,
and discharge air from the balloon in
question.
Clear the buzzer using the STOP switch
and check the tube connections. Connect
the tube if it is about to come off.
Clear the buzzer using the STOP switch
and discharge air from the balloon, and
remove the endoscope and the over-tube.
Replace the scope balloon or the
over-tube.
Clear the buzzer using the STOP switch
and check the tube connections. Connect
the tube if it is about to come off.
Clear the buzzer using the STOP switch.
Turn the power off. After at least five
seconds, turn the power back to on.
If the same error occurs again, remove the
tube from the air inlet of the endoscope
and over-tube and carry out evacuation
manually using a syringe. Then, withdraw
the endoscope and over-tube while
making sure of safety on the monitor
screen and stop the examination.
Turn the power off. After at least five
seconds, turn the power back to on.
If the same error occurs again, remove the
tube from the air inlet of the endoscope
and over-tube and carry out evacuation
manually using a syringe. Then, withdraw
the endoscope and over-tube while
making sure of safety on the monitor
screen and stop the examination.