FujiFilm PB-30 Operation Manual page 69

Balloon controller
Table of Contents

Advertisement

Air supply to the balloon
or air discharge from
the balloon cannot be
stopped.
The pressure indicator
becomes a red display
momentarily during air
feed.
Software malfunction
By the timing of air feed and
pressure display, it is not abnormal.
Remove the tube from the air inlet of the endoscope
and over-tube and carry out evacuation manually
using a syringe. Then, withdraw the endoscope
and over-tube while making sure of safety on the
monitor screen and stop the examination.
This product can be used continuously.
Chapter 12 Troubleshooting
69