(7) Remove the tube kit from the cleaning fluid.
(8) Inject air with a syringe into the tube kit to purge the
cleaning fluid.
(9) Fill a basin with potable water to completely immerse the
tube kit.
(10) Soak the tube kit in the potable water, and rinse it while
gently shaking.
(11) Flush potable water into the tube kit to rinse inside of it.
(12) Repeat steps (10) and (11) two more times for a total of 3
rinses.
(13) Inject potable water into all openings, internal areas and
gaps of the tube kit with a syringe in the potable water to
remove all air bubbles completely.
(14) Pull the tube kit out of the potable water.
(15) To more forcefully expel residual water, it is recommended
to inject filtered compressed air up to 228 kPa for at least
30 seconds into the tube kit to ensure more thorough
drying.
[Note]
It is also recommended to loosely place a sterile gauze
over the opposite (open) end of the tube kit while injecting
compressed air to prevent aerosolization of residual water.
(16) Wipe off the remaining water on the tube kit with a clean
lint-free cloth.
If you find any debris, repeat these procedures.
(17) Dry the tube thoroughly.