FujiFilm PB-30 Operation Manual page 76

Balloon controller
Table of Contents

Advertisement

Main Specifications
Electromagnetic immunity compliance information and guidance
Immunity test
Electrostatic discharge (ESD)
EN 61000-4-2
Electrical fast transient/burst
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
EN 61000-4-11
Power frequency (50/60 Hz)
magnetic field
EN 61000-4-8
76
EN 60601-1-2
Test level
± 6kV: contact
± 8kV: air
± 2kV: for power supply lines
± 1kV: for input/output lines
± 1kV: Line to line
± 2kV: Line to earth
< 11.5V (> 218.5V dip)
For 0.5 cycle
92V (138V dip)
For 5 cycle
161V (69V dip)
For 25 cycle
< 11.5V (> 218.5V dip)
For 5 sec
3 A/m
Compliance
level
Floors should be wood, concrete,
or ceramic tile. If floors are
Same as left
covered with synthetic material,
the relative humidity should be at
least 30%.
Main power quality should be that
Same as left
of a typical commercial or hospital.
Main power quality should be that
Same as left
of a typical commercial or hospital.
Main power quality should be
that of a typical commercial or
hospital. If the user of this product
requires continued operation
during power mains interruptions,
Same as left
it is recommended that this product
is powered from an uninterruptible
power supply or battery.
It is recommended to use this
product by maintaining enough
Same as left
distance from any equipment that
operates with high current.
Guidance