1. Precautions in Using the Product
1) Preparation and inspection before use
Prior to using this product, prepare a spare one to avoid unexpected accidents such as
equipment failure. If a replacement is not available, you may not be able to continue endoscopic
procedures.
Make sure to inspect the equipment before use according to the procedures provided in this manual,
to avoid unexpected accidents, and take full advantage of the equipment's capabilities.
If the inspection result shows any abnormality, do not use the same equipment.
2) Combination of equipment
This product is used in combination with an endoscope system. In order to avoid unexpected
accidents, do not use this product except in combination with components or systems described
in Chapter 2.
Refer to "<Devices Used in Combination>" in "Main Specifications" for the combinations of
endoscopes, over-tubes and tube kits.
In order to avoid unexpected accidents and take full advantage of the equipment's capabilities,
read the operation manual of the equipment carefully before using it.
3) Abnormality in use
If any abnormality is noticed during use, carry out safety checks and discontinue use
immediately.
4) Maintenance
The equipment will wear out and degrade after repeated use for a long period. In particular,
resin and rubber parts wear out over time or due to the chemicals used. Have it checked by
specialists once every six months or once every 100 cases, whichever comes first. Also have
it checked if there is anything wrong with the equipment.
Never disassemble or modify the equipment.