9.3 Chemical Disinfection of Balloon

(1) Fill a basin with a chemical solution prepared according
to chemical solution manufacturer's recommendations to
completely immerse the balloon.
[Note]
When carrying out disinfection, follow the conditions
indicated in "9.2 Conditions of Disinfection."
(2) Completely immerse the balloon in the chemical solution.
(3) Wipe the surface of balloon with sterile gauze in the
chemical solution for at least 15 seconds.
(4) Clean the inside of the balloon by kneading it with fingers
for at least 15 seconds.
(5) Flush the chemical solution with a syringe into all
openings, internal areas and gaps of the balloon in the
chemical solution to remove air bubbles completely.
(6) Flush 90 mL of chemical solution forcefully with a syringe
into internal of balloon in the chemical solution.
(7) Confirm no air bubbles inside and the outer of the balloon.
The balloon must be completely immersed in a chemical solution.
Remove air bubbles completely. If any air bubbles remain, effective
disinfection cannot be achieved and an inadequately balloon may pose
an infection risk.
WARNING
Chapter 9 Chemical Disinfection