Inspecting Pb-30 - FujiFilm PB-30 Operation Manual

Balloon controller
Table of Contents

Advertisement

Chapter 6 Preparation and Inspection

6.5 Inspecting PB-30

6.5.1 Inspecting the Functions of the Pump and the Balloon
(1) Turn on the power switch.
The power switch lights up and self-diagnosis starts.
If there is no abnormality, the balloon switch lights up
in green and air is discharged from the channel on the
over-tube side and that on the endoscope side.
[Note]
If there is any abnormality, an error message will be
displayed on the pressure indicator.
Chapter 12 "Troubleshooting"
(2) Make sure that the inside light is on and the outside light
is off on the toggle switches 1 and 2 of the remote switch.
(3) Press the toggle switch 1 of the remote switch.
Make sure that the outside light lights up and the balloon
on the over-tube inflates.
38