Clean and sterilize (or disinfect) this product properly according to the procedures described
in this manual, immediately after an examination is completed. There is a risk of infection
and/or damaging this product.
The balloon must be completely immersed in chemical solution. Remove air bubbles
completely. If any air bubbles remain, effective disinfection cannot be achieved and an
inadequately disinfected balloon may pose an infection risk.
The tube kit must be completely immersed in chemical solution. Remove air bubbles
completely. If any air bubbles remain, effective disinfection cannot be achieved and an
inadequately disinfected tube kit may pose an infection risk.
Proceed with gas sterilization after removing moisture from the balloon. Applying gas
sterilization to wet surfaces may compromise sterilization.
Proceed with aeration after gas sterilization. Gas residuals remaining in the balloon after gas
sterilization may be harmful to the human body.
Dry the tube kit before storing them. If the tube kit is stored without drying it, there is a risk
of infection.