FujiFilm PB-30 Operation Manual page 35

Balloon controller
Table of Contents

Advertisement

Do not close the vent holes. Otherwise, the instrument may get
overheated.
Install this product in a flat place where there is no vibration or shock. If
not installed, it may fall down or be broken.
Use the power source at the rated voltage. Only use the power cord
provided with this product. Not doing so can create a risk of damage to
the equipment.
Configure the ME system according to the maximum allowable current
of the power strip to be used.
When operating the equipment, wear rubber gloves. Otherwise, the
equipment may malfunction due to static electricity.
Do not put anything on the main unit. Doing so may damage the main
unit.
(1) Place this product attached with filter and tube on the
movable shelf of the cart, so that the pressure gauge can
be observed.
(2) Make sure that the power switch is turned off.
(3) Wrap the remote switch with a waterproof sheet.
(4) Insert the power cord into the power supply connector on
the rear panel of the main unit and connect it securely.
[Note]
Even when the power cord is already connected, make
sure that the power cord is connected securely.
CAUTION
Chapter 6 Preparation and Inspection