6.1.4.6

Preliminary cleaning of the mirror sleeve

Clean the scanner immediately after use as follows:
1. Carefully wipe the mirror sleeve while it is on the scanner so that no
dirt whatsoever can remain stuck and can harden on the surface of
the mirror sleeve. For this purpose, use the following:
- a dampened wiping cloth (see "Cleaning agents  [→ 63]),
- an absorbent cotton gauze or a lint-free cloth, dipped in cleaning
agent (see "Cleaning agents  [→ 63]).
2. Wipe down the mirror sleeve afterwards again with the absorbent
cotton gauze dipped in drinking water.
3. Next dry the mirror sleeve using a lint-free cloth.
6.1.4.7

Disinfecting the scanner body and the mirror sleeve

➢ Use a new cotton gauze moistened with one of the wiping
disinfectants listed in the „Wiping Disinfectants  [→ 63]" section.
Wipe the scanner body first and then the sleeve. This step
disinfects the scanner body and the sleeve. Wipe down the sleeve
afterwards again with the absorbent cotton gauze dipped in drinking
water. Then dry the sleeve with a clean lint-free cloth.
6 Maintenance
6.1 Cleaning, disinfection and sterilization
69