10 Appendix
10.2 Service routines
82
10.2
Service routines
10.2.1
Operation
1. Switch the unit on.
2. Press the Film key, the Sensor key and the Bite Wing key
simultaneously.
The service routine "S01" is displayed in front of a white
background.
3. Scroll through the list of service routines by pressing the +/- keys.
4. Press the Film key to show the setting of the displayed service routine
and make any changes which then may be necessary.
The setting is displayed.
5. To change the setting, press the +/- keys.
6. To save the current service routine, press the Adult key.
7. To discard the settings of the current service routine and quit it, press
the Child key.
8. Finally, switch the unit off and then on again.
Dentsply Sirona
Installation Instructions Heliodent Plus
62 14 972 D 3507
D 3507.031.01.15.02 03.2018