3
Testing U4164A Performance

Equipment Required for the Performance Test

The following equipment is required for the performance test procedure.
Table 1
Equipment
Pulse Generator
150 ps Transition Time Converter (Qty 4)
Flying Lead Probe Set (Qty 2)
SMA/Flying Lead test connectors, (f) SMA to (f) SMA
to Flying Lead Probe (Qty 4)
22
Equipment Required
Critical Specification
≥1275 MHz, two channels, differential outputs,
150-180 ps rise/fall time (if faster, use transition
time converters)
Required if pulse generator's rise time is less than
150 ps. (Pulse generator conditions: Voffset=1V,
δV=250 mV.) Required for 81134A or 8133A opt.
003.
A combination of U4201A & E5382A can be used.
no substitute
Recommended Model/Part
Keysight 81134A or equivalent
Keysight 15435A
Keysight U4203A
See
"Assemble the SMA/Flying Lead Test
Connectors"
on page 23