Design; Technical Support; Safety Instructions; Explanation Of Symbols - Otto Bock E-MAG Active 17B202N Mounting And Service Instructions

Table of Contents

Advertisement

Available languages

Available languages

Medial support 17B205=* left or right - steel
605P8=* Joint Bar System for Medial Support,
light metal in the dimensions 5x20x2000mm
and 5x20x305mm (thickness x width x length)
INFORMATION
The conversion of the flexion components to 7.5° or 10° may only be completed by an authorised
Ottobock Service Centre.

1.6.5 Design

The knee joint is placed 6 mm posteriorly (it is essential to observe recommendations for knee
joint alignment in Section 3). This facilitates easy extension of the knee joint and assures problem-
free unlocking.
The joint bushing (radial bearing) and bearing washer (axial bearing) are made of maintenance-
free plastic. The wedge lock is connected to an electronically controlled spring, which locks it to
the lower joint section. The electronic controller, consisting of a gyroscope and 2D acceleration
sensors to sense the gait phase, measures the joint's position during the gait cycle and opens
the joint accordingly before toe lift-off, locking it again in the extended position before heel strike.
The enclosed joint bar cases serve for attachment of the system joint bar or lamination lancets.
The electronic components are designed so that they can be replaced by a technician. The
batteries can be removed from the lamination frame and charged in the charger. The charger
has two ports so that two batteries can be charged simultaneously.

1.6.6 Technical support

Your national Ottobock Team will be pleased to answer your technical questions. Contact
addresses and telephone numbers are listed on the last page of this manual.

2 Safety instructions

2.1 Explanation of symbols

Warnings regarding possible risks of accident or injury.
CAUTION
Warnings regarding possible technical damage.
NOTICE
Additional information on the fitting/use.
INFORMATION

2.2 General safety instructions

INFORMATION
Certification for the E-MAG Active is mandatory. Without an orthotist certified by name, the E-MAG
Active is not CE compliant and therefore CE certification expires from an industrial perspective.
44 | Ottobock
E-MAG Active Lower Flexion Components (in
5°, 7.5° and 10°)
E-MAG Active