CAUTION
Risk of damage due to improper installation. Please ensure that no notches are made in the splint
material during milling since this creates predetermined breaking points.
Carefully cut a cable groove into the joint guide (Figure 22/23) and then reinsert the knee joint
dummy.
Measure the length of the cable dummies included
• Dummy for the cable between the knee joint and electronics (7 mm wide, Figure 24).
• Perlon wire for the cable connecting the electronics to the battery (Figure 25).
Fasten the cable dummies into the designated recesses of the
tery as follows:
• From the knee joint to the electronics (Figure 26).
• From the electronics to the battery (Figure 27).
52 | Ottobock
22
:
24
electronics and bat-
dummies for
E-MAG Active
23
25