Components Of The Supply Module; Service Block; Change The Water And Air Filters - Dentsply Sirona Sim Multi Bench Operating Instructions Manual

Simulation unit
Hide thumbs Also See for Sim Multi Bench:
Table of Contents

Advertisement

5 Care, cleaning and maintenance by the user

5.5 Components of the supply module

A
B
C
94
Components of the supply module
5.5
5.5.1

Service block

5.5.1.1

Change the water and air filters

If you notice any changes in media flows, check the water and air filters
for permeability. The filters must be changed, if necessary.
To reorder the filters, see "Spare parts and consumables".
1. Switch the simulation unit off at the power switch.
Ä The water and air supply are switched off.
2. Remove side cover A and loosen the connection for the protective
ground wire B.
B
3. Any remaining water flows out of the water filter when it is opened.
You should therefore place an absorbent cloth underneath the filter.
Then unscrew and remove the screw cap of the water filter B and/or
the air filter C.
4. Check the filters and replace them if necessary.
Operating Instructions Simulation Unit Sim Multi Bench
D3684.201.01.02.02    07.2019
Dentsply Sirona
66 82 384 D3684