Command
FORever
TIME
<time> A real number: 0.1 to 1440.0
WAVeforms WAVeforms sets the maximum number of waveforms that are required before
<number_of_
waveforms>
Example

RUMode

:MTESt:RUMode {FORever | TIME, <time> | WAVeforms,
<number_of_waveforms>}
The :MTESt:RUMode command determines the termination conditions for the
mask test. The choices are FORever, TIME, or WAVeforms.
If WAVeforms is selected, a second parameter is required indicating the number
of failures that can occur or the number of samples or waveforms that are to be
acquired.
FORever runs the Mask Test until the test is turned off. This is used when you
want a measurement to run continually and not to stop after a fixed number of
failures. For example, you may want the Mask Test to run overnight and not
be limited by a number of failures.
TIME sets the amount of time in minutes that a mask test will run before it
terminates.
the mask test terminates.
An integer: 1 to 1,000,000,000.
This example sets the mask test subsystem run until mode to continue testing
until 500,000 waveforms have been gathered.
10 OUTPUT 707;"MTEST:RUMODE WAVEFORMS,500E3"
20 END
Mask Test Commands
RUMode
22-33