Set the Mode Before Executing Commands
Before you can execute the :TRIGger:ADVanced:STATe commands, set
the mode by entering:
:TRIGger:MODE ADVanced
and
:TRIGger:ADVanced:MODE STATe
To query the oscilloscope for the advanced trigger mode, enter:
:TRIGger:ADVanced:MODE?
The :TRIGger:ADVanced:STATe commands define the conditions for
the State Trigger Mode. As described in the following commands, you
set up the :TRIGger:ADVanced:STATe commands with the following
commands and queries:
• CLOCk
• LOGic
• LTYPe
• SLOPe
• THReshold
29-39