Table 7-8
All Other Models Points Value Ranges for Single Sweep Trigger Mode
Memory Option
Standard
Half Channel Mode
Full Channel Mode
040 Option
Half Channel Mode
Full Channel Mode
080 Option
Half Channel Mode
Full Channel Mode
160 Option
Half Channel Mode
Full Channel Mode
320 Option
Half Channel Mode
Full Channel Mode
640 Option
Half Channel Mode
Full Channel Mode
Equivalent Time
Sampling Mode for All
Models
Example
Below Maximum Real Time Sampling Mode
Normal
Peak Detect
16 to 2050000
16 to 512500
16 to 2050000
16 to 512500
16 to 4100000
16 to 1025000
16 to 4100000
16 to 1025000
16 to 8200000
16 to 2050000
16 to 8200000
16 to 2050000
16 to 16400000
16 to 4100000
16 to 16400000
16 to 4100000
16 to 32800000
16 to 8200000
16 to 3280000
16 to 8200000
16 to 65600000
16 to 16400000
16 to 65600000
16 to 16400000
Equivalent Time Sampling mode takes the oscilloscope out of half channel mode
and the memory depth range is 16 points to 32768 points.
Interaction between :ACQuire:SRATe and :ACQuire:POINts
If you assign a sample rate value with :ACQuire:SRATe or a points value using
:ACQuire:POINts the following interactions will occur. "Manual" means you are
setting a non-AUTO value for SRATe or POINts.
SRATe
POINts
Result
AUTO
Manual
POINts value takes precedence (sample rate is limited)
Manual
AUTO
SRATe value takes precedence (memory depth is limited)
Manual
Manual
SRATe value takes precedence (memory depth is limited)
This example sets the memory depth to 500 points.
10
OUTPUT 707;":ACQUIRE:POINTS 500"
20
END
High Resolution
Averaging
16 to 1025000
2097152
16 to 1025000
1048576
16 to 2050000
2097152
16 to 2050000
1048576
16 to 4100000
2097152
16 to 4100000
1048576
16 to 8200000
2097152
16 to 8200000
1048576
16 to 16400000
2097152
16 to 16400000
1048576
16 to 32800000
2097152
16 to 32800000
1048576
Acquire Commands
POINts
7-19