Disk Commands
CSV, TSV and TXT Header Format
XDispOrg
The number of X display origin columns (n)
depends on the number of sources being stored.
The X display origin is the X-axis value at the left
edge of the display. For time domain waveforms,
it is the time at the start of the display. This value
is treated as a double precision 64-bit floating
point number. If the value is zero then no data
has been acquired.
XInc
The number of X increment columns (n)
depends on the number of sources being store.
The X increment is the duration between data
points on the X axis. For time domain
waveforms, this is the time between points. If
the value is zero then no data has been acquired.
XOrg
The number of X origin columns (n) depends on
the number of sources being store. The X origin
is the X-axis value of the first data point in the
data record. For time domain waveforms, it is
the time of the first point. This value is treated
as a double precision 64-bit floating point
number. If the value is zero then no data has
been acquired.
XUnits
The number of X units columns (n) depends on
the number of sources being store. The X units
is the unit of measure for each time value of the
acquired data.
YDispRange
The number of Y display range columns (n)
depends on the number of sources being store.
The Y display range is the Y-axis duration of the
waveform which is displayed. For voltage
waveforms, it is the amount of voltage across the
display. If the value is zero then no data has been
acquired.
YDispOrg
The number of Y display origin columns (n)
depends on the number of sources being store.
The Y-display origin is the Y-axis value at the
center of the display. For voltage waveforms, it
is the voltage at the center of the display. If the
value is zero then no data has been acquired.
13-20