Preamble Definition
<revision>
<preamble_
<type>
data>
<start>
<points>
<count 1...n>
<X display range 1...n>
<X display origin 1...n>
<X increment 1...n>
<X origin 1...n>
<X units 1...n>
<Y display range 1...n>
<Y display origin 1...n>
<Y increment 1...n>
<Y origin 1...n>
<Y units 1...n>
<frame model #>
<date>
<time>
<max bandwidth limit>
<min bandwidth limit>
<revision> 0
Always zero.
<type> RAW type.
AVERage type.
VHIStogram.
HHIStogram.
INTerpolate.
DIGital.
PDETect.
<start> 0
Always zero.
<points> The number of data points or data pairs contained in the waveform data.
<count 1...n> The number of count columns (n) depends on the number of sources being
stored. For the AVERAGE waveform type, the count value is the fewest number
of hits for all time buckets. This value may be less than or equal to the value
requested with the :ACQuire:AVERage:COUNt command. For NORMAL, RAW,
and INTerpolate this value is 0 or 1.
Disk Commands