Agilent Technologies Infiniium 8000A Programmer's Reference Manual page 872

Table of Contents

Advertisement

Element
Description
X Display Origin
The X display origin is the X-axis value at the left edge of the display.
For time domain waveforms, it is the time at the start of the display.
This value is treated as a double precision 64-bit floating point number.
(See the :WAVeform:XDISplay? query.)
Y Display Range
The Y display range is the Y-axis duration of the waveform which is displayed.
For voltage waveforms, it is the amount of voltage across the display. (See
the :WAVeform:YRANge? query.)
Y Display Origin
The Y-display origin is the Y-axis value at the center of the display.
For voltage waveforms, it is the voltage at the center of the display.
(See the :WAVeform:YDISplay? query.)
Date
The date that the waveform was acquired or created.
Time
The time that the waveform was acquired or created.
Frame Model #
The model number of the frame that acquired or created this waveform.
The frame model number is ignored when it is sent to an oscilloscope in the
preamble.
Acquisition Mode
The acquisition sampling mode of the waveform. (See :ACQuire:MODE.)
Complete
The complete value is the percent of time buckets that are complete. The
complete value is ignored when it is sent to the oscilloscope in the preamble.
(See the :WAVeform:COMPlete? query.)
X Units
The X-axis units of the waveform. (See the :WAVeform:XUNits? query.)
Y Units
The Y-axis units of the waveform. (See the :WAVeform:YUNits? query.)
Band Pass
The band pass consists of two values that are an estimation of the maximum
and minimum bandwidth limits of the source waveform. The bandwidth limit
is computed as a function of the selected coupling and filter mode.
(See the :WAVeform:BANDpass? query.)
See Also
Waveform Commands
PREamble
:WAVeform:DATA?
30-50

Advertisement

Table of Contents
loading

Table of Contents