Agilent Technologies Infiniium 8000A Programmer's Reference Manual page 243

Table of Contents

Advertisement

Table 11-5
Status Byte Register Bits
Bit
Bit Weight
Bit Name
7
128
OPER
6
64
RQS/MSS
5
32
ESB
4
16
MAV
3
8
---
2
4
MSG
1
2
USR
0
1
TRG
0 = False = Low
Condition
0 = no enabled operation status conditions have occurred
1 = an enabled operation status condition has occurred
0 = oscilloscope has no reason for service
1 = oscilloscope is requesting service
0 = no event status conditions have occurred
1 = an enabled event status condition has occurred
0 = no output messages are ready
1 = an output message is ready
0 = not used
0 = no message has been displayed
1 = message has been displayed
0 = no enabled user event conditions have occurred
1 = an enabled user event condition has occurred
0 = no trigger has occurred
1 = a trigger occurred
1 = True = High
Common Commands
*STB? (Status Byte)
11-21

Advertisement

Table of Contents
loading

Table of Contents