Disk Commands
BIN Header Format
This value is one except for peak detect data and
digital data.
Count
An integer (4 byte signed) which is the number
of hits at each time bucket in the waveform
record when the waveform was created using an
acquisition mode like averaging. For example,
when averaging, a count of four would mean
every waveform data point in the waveform
record has been averaged at least four times.
The default value is 0.
X Display Range
A float (4 bytes) which is the X-axis duration of
the waveform that is displayed. For time domain
waveforms, it is the duration of time across the
display. If the value is zero then no data has
been acquired.
X Display Origin
A double (8 bytes) which is the X-axis value at
the left edge of the display. For time domain
waveforms, it is the time at the start of the
display. This value is treated as a double
precision 64-bit floating point number. If the
value is zero then no data has been acquired.
X Increment
A double (8 bytes) which is the duration
between data points on the X axis. For time
domain waveforms, this is the time between
points. If the value is zero then no data has been
acquired.
X Origin
A double (8 bytes) which is the X-axis value of
the first data point in the data record. For time
domain waveforms, it is the time of the first
point. This value is treated as a double precision
64-bit floating point number. If the value is zero
then no data has been acquired.