Table 11-2
See Also
Common Commands
*ESE (Event Status Enable)
The Standard Event Status Enable Register contains a mask value for the bits
to be enabled in the Standard Event Status Register. A "1" in the Standard Event
Status Enable Register enables the corresponding bit in the Standard Event
Status Register. A "0" in the enable register disables the corresponding bit.
Standard Event Status Enable Register Bits
Bit
Weight
Enables
7
128
PON - Power On
6
64
5
32
CME - Command Error
4
16
EXE - Execution Error
3
8
DDE - Device Dependent Error
2
4
QYE - Query Error
1
2
RQC - Request Control
0
1
OPC - Operation Complete
Refer to the chapter, "Status Reporting," for a complete discussion of status.
11-6
Definition
Indicates power is turned on.
Not Used.
Permanently set to zero.
Indicates whether the parser detected an
error.
Indicates whether a parameter was out of
range, or was inconsistent with the current
settings.
Indicates whether the device was unable to
complete an operation for device-
dependent reasons.
Indicates if the protocol for queries has been
violated.
Indicates whether the device is requesting
control.
Indicates whether the device has completed
all pending operations.