CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL. Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent products. For any assistance, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-5. Printing Copies of Documentation from the Web...
Page 4
Introduction ............... . 6-2 A. Standard Definitions Class Assignments and Standard Definitions Values are Available on the Web ....A-2 85054D...
Calibration Kit Overview Calibration Kit Overview The Agilent 85054D type-N calibration kit is used to calibrate Agilent network analyzers up to 18 GHz for measurements of components with 50Ω type-N connectors. The standards in this calibration kit allow you to perform simple 1- or 2-port and TRM (thru–reflect–match) calibrations.
Equipment Required but Not Supplied Connector cleaning supplies and various electrostatic discharge (ESD) protection devices are not supplied with the calibration kit but are required to ensure successful operation of the kit. Refer to Table 6-2 on page 6-3 for ordering information. 85054D...
“Contacting Agilent” on page 5-5. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. .“Returning a Kit or Device to Agilent” on page 5-4. 85054D...
Record these serial numbers in Table 1-1. Recording the serial numbers will prevent confusing the devices in this kit with similar devices from other kits. Table 1-1 Serial Number Record for the 85054D Device Serial Number Calibration kit _______________________________...
Calibration Kits Documented in This Manual Calibration Kits Documented in This Manual This manual applies to any 85054D calibration kit whose serial number prefix is listed on the title page. If your calibration kit has a different serial number prefix, refer to the “Calibration Kit History”...
With PSCs on test ports and standards, the percentage of accuracy achieved when measuring at 50 dB return loss levels is comparable to using conventional slotted connectors measuring devices having only 30 dB return loss. This represents an accuracy improvement of about 10 times. 85054D...
Improper connections, resulting from pin depth values being out of the observed limits (see Table 2-2 on page 2-4), or from bad connections, can also damage these devices. 85054D...
Periodically remeasure the device and note any changes in its corrected response which can be attributed to the test system. With experience you will be able to see changes in the reference responses that indicate a need to perform the measurement calibration again. 85054D...
Page 14
General Information When to Calibrate 1-10 85054D...
During a measurement calibration, the temperature of the calibration devices must be stable and within the range specified in Table 2-1. Avoid unnecessary handling of the devices during calibration because your IMPORTANT fingers are a heat source. 85054D...
Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to the electrical performance of devices. Agilent Technologies verifies the mechanical characteristics of the devices in this kit with special gaging processes and electrical testing.
Page 18
When measuring pin depth, the measured value (resultant average of three or more measurements) is not the true value. Always compare the measured value with the observed pin depth limits in Table 2-2 to evaluate the condition of device connectors. 85054D...
A-2). Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities...
3. Connect the other end of the cable to the test port. 4. Remove the short from the cable. Refer to Chapter 6 , “Replaceable Parts,” for part numbers and instructions for ordering ESD protection devices. Figure 3-1 ESD Protection Setup 85054D...
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration. 85054D...
It may be harmful if absorbed through the skin. Wash thoroughly after handling. In case of spill, soak up with sand or earth. Flush spill area with water. Dispose of isopropyl alcohol in accordance with all applicable federal, state, and local environmental regulations. 85054D...
Page 25
Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 4. Inspect Inspect the connector again to make sure that no particles or residue are present. 85054D...
Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Connectors The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. (See Table 6-1 on page 6-2 for part number information.) They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.
After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis. e. Compare the average reading with the observed pin depth limits in Table 2-2 on page 2-4. 85054D...
Page 29
Use, Maintenance, and Care of the Devices Gaging Connectors Figure 3-3 Gaging Male Type-N Connectors 85054D...
Page 30
Compare the average reading with the observed pin depth limits in Table 2-2 on page 2-4. 3-10 85054D...
Page 31
Use, Maintenance, and Care of the Devices Gaging Connectors Figure 3-4 Gaging Female Type-N Connectors 85054D 3-11...
A torque wrench is not included in the calibration kit. Refer to Chapter 6 for part number and ordering information. Table 3-1 Torque Wrench Information Connector Type Torque Setting Torque Tolerance ±13.5 N-cm (±1.2 in-lb) Type-N 135 N-cm (12 in-lb) 3-12 85054D...
Page 33
This is especially true when several devices are connected together. Figure 3-5 Wrench Positions 2. Hold the torque wrench lightly, at the end of the handle only (beyond the groove). See Figure 3-6. Figure 3-6 Using the Torque Wrench 85054D 3-13...
• Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3-14 85054D...
These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in this kit is traceable to NIST through Agilent Technologies. 85054D...
A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of this kit. For more information, contact Agilent Technologies. For contact information, see page 5-5.
Troubleshooting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1. Figure 5-1 Troubleshooting Flowchart 85054D...
2. Enter your VNA model number (Ex: 8753ES) in the Search box and click Search. 3. Under the heading Manuals & Guides, click on the title/hyperlink for the document PDF you want to view. If you need additional information, see “Contacting Agilent” on page 5. 85054D...
Troubleshooting Returning a Kit or Device to Agilent Returning a Kit or Device to Agilent If your kit or device requires service, contact Agilent Technologies for information on where to send it. See “Contacting Agilent” on page 5-5 for contact information. Include a...
If you do not have access to the Internet, please contact your Agilent field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period. 85054D...
Replaceable Parts Introduction Introduction Table 6-1 lists the replacement part numbers for items included in the 85054D calibration kit and Figure 6-1 illustrates each of these items. Table 6-2 lists the replacement part numbers for items recommended or required for successful operation but not included in the calibration kit.
Page 47
Foam tipped cleaning swabs 9301-1243 Refer to “Clarifying the Terminology of a Connector Interface” on page 1-8. b. Agilent can no longer safely ship isopropyl alcohol, so customers should purchase it locally. Figure 6-1 Replaceable Parts for the 85054D Calibration Kit 85054D...
Agilent vector network analyzers. It also provides some examples of how to set up a new calibration kit and how to modify an existing calibration kit definition file. To download a free copy, go to www.agilent.com and enter literature number 5989-4840EN in the Search window. 85054D...
Page 51
3-10 entering connector gage accuracy calibration kit constants, calibration, See calibration definitions Agilent Application Note contacting Agilent Technologies contents contents history incomplete modifying definition files what to do overview calibration report calibration sticker...
Page 52
3-14 parts how often to calibrate ordering humidity replaceable required but not supplied performance verification incoming inspection fail inspection pin depth damage definition of Index-2 85054D...
Page 53
3-10 humidity zeroing connector gage 3-10 pin depth return loss temperature verifying standard definitions downloading from Agilent Web site standards calibration NIST static discharge sticker calibration storage 3-14 temperature supplemental characteristics supplies cleaning Index-3 85054D...