1
Calibration Procedures
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed below in
available, substitute with another calibration standard of equivalent accuracy.
Table 1-1
Recommended test equipment
Application
DC voltage
DC current
Resistance
AC voltage
AC current
Frequency
Capacitance
Diode
Temperature
Short
20
Recommended equipment
Fluke 5520A
Fluke 5520A and
Fluke 5500A/COIL
Fluke 5520A
Fluke 5520A
Fluke 5520A and
Fluke 5500A/COIL
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Shorting plug — a dual banana plug with a copper wire shorting the two terminals
Table
1-1. If the exact instrument is not
Recommended accuracy requirement
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
<20% of the instrument accuracy specification
Keysight U1190A Series Service Guide