Operating Instruction EXTRAmatic 9040
6 Preparation methods
|
6.2 Maintenance
▶ Fit the hood ③ (bevel towards the front), press on and lock with the two snap-
locks.
2
3
Further assembly.
See also: 6.2.1 Installation, Page 50
Note
During active carbon filter replacement, a reset must be performed once new active
carbon filters have been installed (as per utility program: "Reset active carbon filter
monitoring time").
See also: 5.3.13 "Reset active carbon filter monitoring time" utility program (2550),
Page 45
At the same time, the "change time" can be adjusted to between 100 and 2,500
operating hours as required (as per utility program: "Time to change active carbon
filter").
See also: 5.3.11 "Time to change active carbon filter" utility program (2450),
Page 45
6.2.3 Replacing the long-term filter
See also:
7 Troubleshooting, Page 59
1
53/62