Advertisement

4. Please confirm if the values of Meter, Load & DVM meet to that in Table 10-7.
Source Setting
Frequency
(Hz)
45/440

10.5 DC Performance Test

10.5.1 I monitor Measurement Verification

DC Source wiring as Figure 10-2, switch AC Load to DC mode, the procedures are as follows.
1. Press CONF./LOCAL.
2. Press 1 and ENTER to enter SETUP.
C H O O S E = 1
3. Move the cursors of ▲ and ▼ to "1-1. MODE SELECT".
4. Press direction keys of ◄, ► or turn rotary knob to select "DC" and then press ENTER.
1 - 1 . M O D E
1 - 2 . S H O R T
1 - 3 . L O C K
1 - 4 . S O U N D
1 - 5 . C F / P F
1 - 6 . T I M I N G
1 - 7 . E X T E R N A L
1 - 8 . A B A
Table 10-7
Meter & Load
Measurement
Voltage
V Specification
(V)
10V
9.64~10.36
350V
349.30~350.70
1 . S E T U P
2 . R E M O T E
3 . C A L I B R A T I O N
4 . D I S P L A Y
5 . F A C T O R Y
6 . M O D E L
7 . P A R A L L E L
S E L E C T
M O D E
S E T T I N G = B O T H
M O D E
M O D E
V_Mon
Specification
0.179~0.221
6.945~7.055
S E L E C T I O N
D E F A U L T
I N F O R M A T I O N
M O D E
= D C
/
A
C
= O F F
/ T O G G L E / H O L D
= R O T A R Y
/ K E Y P A D / A L L
= O N
/ O F F
/ C F / P F
= O F F
/ H O L D U P / T R A N S F E R
= O F F
/ O N
= E N A B L E
/ D I S A B L E
Verification
[ C O N F ]
[ S E T U P ]
10-7