Ac Load Simulation - Chroma 63800 Series Operation & Programming Manual

Programmable ac/dc electronic load
Table of Contents

Advertisement

I s e t = 0 . 5 0 0
I p ( m a x ) = 1 0 . 0
V = 0 . 0 0
I = 0 . 0 0
A C
P R I O R I T Y : C F
If user chooses Lock ALL thus all keypads and rotary are disabled except for
LOCK key.
3.5

AC Load Simulation

The 63800 series electronic load has AC/DC simulated functions. The AC load simulation
includes constant load mode and rectified load mode and its operation modes are described
as below.
In AC mode, the CC, CR, CP, RLC/DC RECT keys are used to select the load modes for
local control. AC mode and DC mode can be chosen in the 63800 series. Press
CONF./LOCAL and select the first item, "SETUP", to choose one. In AC mode, the load
values, PF and CF are common to CC, and CP modes; CR and RLC mode only set
determinate values. It is necessary to set UUT's output maximum current value I
modes. The value I
(max) is for 63800 to test UUT's output impedance for getting system
p
optimum.
The UUT's output meets AC mode and frequency is within 45Hz ~
440Hz, output voltage 45Vrms ~ 350Vrms and output inductance is less
than 2mH thus it can normal use.
63800's power-on default value is AC mode.
If user wants to return to AC mode from DC mode.
1. Press CONF./LOCAL key.
2. Press 1 , then press ENTER to enter SETUP.
C F = 1 . 4 1 4
P = 0 . 0 0
S = 0 . 0 0
Local Operation
P F = 1 . 0 0 0
U N I T : A
(max) in all
p
3-13

Advertisement

Table of Contents
loading