3.17.3 CP Mode
Under timing mode to enter CP timing mode press the CP key.
In this mode the loading power (P
time and cutoff voltage of the UUT can be set.
P s e t = 1 0 0 . 0 0 W
T I M E O U T = 0 0 : 0 0 : 0 0
V = 0 . 0 0
T
= 0 0 : 0 0 : 0 0 : 0 0 0 m s
H O L D U P
D C
The procedures of setting P
9.5, are described as below:
1. Move the cursor to the command line of "P
P s e t = 1 0 0 . 0 0 W
T I M E O U T = 0 0 : 0 0 : 0 0
2. Press 3 , 0 , 0 then press ENTER to change the value to "300.00".
P s e t = 3 0 0 . 0 0 W
T I M E O U T = 0 0 : 0 0 : 0 0
3. Move the cursor by pressing ( ◄, ► ) to the command line of "I
P s e t = 3 0 0 . 0 0 W
T I M E O U T = 0 0 : 0 0 : 0 0
4. Press 1 , 6 then press ENTER to change the value to "16.0".
P s e t = 3 0 0 . 0 0 W
T I M E O U T = 0 0 : 0 0 : 0 0
5. Move the cursor by pressing ( ◄, ► ) to the command line of "TIMEOUT=00:00:00".
), maximum current value I
set
I p ( m a x ) = 5 0 . 0 A
C U T O F F
= 300.00, I
(max) = 16.0, Timeout = 00:10:15, and Cutoff V =
set
p
=".
set
(max), as well as the timeout
p
V = 5 0 . 0 0 0
I p ( m a x ) = 5 0 . 0 A
C U T O F F
V = 5 0 . 0 0 0
I p ( m a x ) = 5 0 . 0 A
C U T O F F
V = 5 0 . 0 0 0
(max)=".
p
I p ( m a x ) = 5 0 . 0 A
C U T O F F
V = 5 0 . 0 0 0
I p ( m a x ) = 1 6 . 0 A
C U T O F F
V = 5 0 . 0 0 0
Local Operation
3-69