Time Subsystem - Chroma 63800 Series Operation & Programming Manual

Programmable ac/dc electronic load
Table of Contents

Advertisement

Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
Description: It sets SYNC mode. When SYNC sets as 1, it indicates the system uses line
sync to do synchronization. When SYNC sets as 0, it indicates the system uses the frequency
users set to do synchronization. This command is only operating in DC Rectified Mode.
Setting Syntax
: [LOAD:]SYNC:DC
Setting Parameter : 0 | OFF, 1 | ON
Query Syntax
: [LOAD:]SYNC:DC?
Response Parameter: 0 | 1
Example
: SYNC:DC ON
[LOAD:]FREQ[:LEVel][:AMPLitude]:DC
Description: It sets frequency. When SYNC sets as 0, it indicates the system uses the
frequency users set to do synchronization. This command is only operating in DC Rectified
Mode.
Setting Syntax
: [LOAD:]FREQ[:LEVel][:AMPLitude]:DC
Setting Parameter : <NR2>, Range: 40 ~ 440, Unit: Hz
Query Syntax
: [LOAD:]FREQ[:LEVel][:AMPLitude]:DC?
Response Parameter: <NR2>
Example
: FREQ:DC 100

8.6.3.10 Time Subsystem

[LOAD:]
TIME
:MODE
:TOUT
:VCUT
[LOAD:]TIME:MODE
Description
: It sets TIMING mode. When the timing mode is set, user can choose
Setting Syntax
: [LOAD:]TIME:MODE
Setting Parameter : 0|OFF,1|HOLD,2|TRAN
Query Syntax
: [LOAD:]TIME:MODE?
Response Parameter: OFF,HOLD,TRAN
Example
: TIME:MODE HOLD
[LOAD:]TIME:TOUT
Description
: It sets the timeout. When the AC load is load on, the timer will not stop
Setting Syntax
: [LOAD:]TIME:TOUT
Setting Parameter : <NR1>, Range: 0 ~ 215999,Unit: Sec
Query Syntax
: [LOAD:]TIME:TOUT?
Response Parameter: <NR1>
Example
: TIME:TOUT 180
[LOAD:]TIME:VCUT
Description
: It sets the cutoff voltage. When the voltage is under the cut off voltage.
Setting Syntax
: [LOAD:]TIME:VCUT
Setting Parameter : <NR2>, Range: 0 ~ 500.000,Unit: V
Query Syntax
: [LOAD:]TIME:VCUT?
Response Parameter: <NR2>
Example
: TIME:VCUT
8-22
Select line frequency or set frequency
HOLD or TRANSFER mode.
counting until meet the setting value of run time set by users.
The timer will stop counting.

Advertisement

Table of Contents
loading

This manual is also suitable for:

638046380263803