Status Registers; Status Subsystem Commands - Chroma 63800 Series Operation & Programming Manual

Programmable ac/dc electronic load
Table of Contents

Advertisement

Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
9.3

Status Registers

The Questionable Status registers inform you one or more questionable status
conditions, which indicate certain errors or faults have occurred. Table 9-1 lists the
questionable status conditions that are applied to the electronic load.
When the corresponding bit of Questionable Status Condition register is set, it indicates
the condition is true.
Reading of the Questionable Status Event register will reset it to zero.
The Questionable status Enable register can be programmed to specify the questionable
status event bit that is logically ORed to become Bit 3 (QUES bit) in the Status Byte
register.
Value Bit
Mnemonic Bit
FE
0
FREQ ERR
1
UV
2
OV
3
OC
5
OP
7
The Operation Status registers inform you one or more operation status conditions,
which indicate certain errors or faults have occurred. Table 9-2 lists the operation status
conditions that are applied to the electronic load.
When the corresponding bit of Operation Status Condition register is set, it indicates the
condition is true.
Reading of the Operation Status Event register will reset it to zero.
The Operation Status Enable register can be programmed to specify the operation status
event bit that is logically ORed to become Bit 7 (OPER bit) in the Status Byte register.
Value Bit
Mnemonic Bit
Weight
LF
0
OT
1
FF
2

9.3.1 Status Subsystem Commands

STATus:
QUEStionable:
EVENt
CONDition
ENABle
STATus:QUEStionable:EVENt?
Description
9-2
Table 9-1 Bit Description of Questionable Status
Weight
1
Under frequency range or over frequency range.
2
Frequency variation occurred condition while loading.
AC Mode voltage is lower than 45V, DC Mode voltage is
4
lower than 7.5V.
8
AC Mode voltage is higher than 350V.
32
AC/DC Mode current is higher than 18A or 45A.
64
AC/DC Mode power is higher than 1800W or 4500W.
Table 9-2 Bit Description of Operation Status
1
Load fail
2
Over temperature protection
4
FAN FAIL
Return the value of the event register
Return the value of the condition register
Enable specific bits in the Event register
:
This query is the value of the Questionable Event register.
The Event register is a read-only register, which holds all
events. Reading the Questionable Event register clears it.
Meaning
Meaning

Advertisement

Table of Contents
loading

This manual is also suitable for:

638046380263803

Table of Contents