6. AC Source settings are shown as Table 10-5. Please follow the model to set W, PF &
I
(max). According to the model to confirm if the values of Meter & Load meet to that in
p
Table 10-5.
Source Setting
Voltage
Frequency
(V)
(Hz)
50
45/440

10.4.5 THD Measurement Verification

AC Source sets output voltage waveform to sine, AC Load enters CC mode and no loading.
Select ▼ at the upper side of rotary knob to press for twice, observe if THD measurement
value of Load and Meter panel meet to Table 10-6.
I s e t = 0 . 5 0 0
I p ( m a x ) = 1 0 . 0
I P+ = 0 . 0 0
I P- = 0 . 0 0
A C
P R I O R I T Y : C F
Source Setting
Voltage
Frequency
(V)
150
10.4.6 V monitor, Voltage Measurement Verification
1. Press CC to enter CC mode.
2. Set AC SOURCE to series file to output current limit 1A, other settings are as Table 10-7.
3. Switch DVM measurement terminal to Front and DVM range to AC for measuring V_Mon
voltage.
10-6
Table 10-5
Load Setting
W
PF
I
p
50
0.2
1800
0.6
50
0.2
3600
0.8
50
0.2
4500
1
C F = 1 . 4 1 4
Table 10-6
Meter & Load Measurement
THD
(Hz)
45
10%
440
40%
Meter & Load Measurement
P
(max)
Specification
44.5~55.5
54
1791~1809
39.1~60.9
108
3582~3618
36.4~63.3
135
4477.5~4522
.5
P F = 1 . 0 0 0
T H D = 0 . 0 0
THD(%) Specification
9.9~10.1
39.6~40.4
PF
Model
Specification
0.19~0.21
63802
0.59~0.61
0.19~0.21
63803
0.79~0.81
0.19~0.21
63804
0.99~1.01
U N I T : A
Model
AC Load