Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
5. Press CC to return to main screen of DC mode.
6. High current DC Source settings are shown as Table 10-8, switch DVM2 measurement
terminal to Rear. Please follow the model under CC mode to set loading current & I
According to the model to confirm if the values of DVM meet to that in Table 10-8.
Source Setting
Voltage
(V)
100
10.5.2 CC Mode Operation & Measurement Verification
1. Press CC to return to main screen of DC CC mode.
2. High current DC Source settings are shown as Table 10-9. Please follow the model under
CC mode to set loading, S.R. & I
Meter & Load meet to that in Table 10-9.
Source Setting
Voltage (V) Current (A) Current (A)
100

10.5.3 CR Mode Operation Verification

1. Press CR to return to main screen of DC CR mode.
2. High current DC Source settings are shown as Table 10-10. Please follow the model
under CR mode to set R value & I
Meter & Load meet to that in Table 10-10.
10-8
Table 10-8
Load Setting
Current
I
(A)
1
18
1
50
36
1
45
(max). According to the model to confirm if the values of
p
Table 10-9
Load Setting
S.R.
5
18
5
50
100
36
5
45
(max). According to the model to confirm if the values of
p
DVM
I_Mon
I
(max)
p
Specification
0.03~0.07
54
0.876~0.925
0.03~0.07
108
1.771~1.829
0.03~0.07
135
2.219~2.281
Meter & Load
Measurement
Ip(max)
I Specification (A)
4.959~5.041
54
17.946~18.054
4.923~5.077
108
35.892~36.108
4.905~5.095
135
44.865~45.135
(max).
p
Model
63802
63803
63804
Model
63802
63803
63804