Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
[LOAD:] RLC:LS
Description: It sets the value Ls, this command is only operating in AC RLC Mode.
Setting Syntax
: [LOAD:] RLC:LS
Setting Parameter : <NR1>, Range: 0 ~ 9999, Unit: uH
Query Syntax
: [LOAD:] RLC:LS?
Response Parameter: <NR1>
Example
: RLC:LS 200
[LOAD:] RLC:RL
Description: It sets the value RL, this command is only operating in AC RLC Mode.
Setting Syntax
: [LOAD:] RLC:RL
Setting Parameter : <NR2>, 63804 Range: 1.11 ~ 9999.99, 63803 Range: 1.39 ~ 9999.99,
Query Syntax
: [LOAD:] RLC:RL?
Response Parameter: <NR2>
Example
: RLC:RL 50
[LOAD:] RLC:RS
Description: It sets the value RS, this command is only operating in AC RLC Mode.
Setting Syntax
: [LOAD:] RLC:RS
Setting Parameter : <NR2>, Range: 0 ~ 9.999, Unit: Ohm
Query Syntax
: [LOAD:] RLC:RS?
Response Parameter: <NR2>
Example
: RLC:RS 0.1
[LOAD:] RLC:POW
Description: It sets the value C, this command is only operating in AC RLC CP Mode.
Setting Syntax
: [LOAD:] RLC:POW
Setting Parameter : <NR2>, Range: 100 ~ 4500, Unit: W
Query Syntax
: [LOAD:] RLC:POW?
Response Parameter: <NR2>
Example
: RLC:POW 1000
[LOAD:] RLC:PFAC
Description: It sets the value PF, this command is only operating in AC RLC CP Mode.
Setting Syntax
: [LOAD:] RLC:PFAC
Setting Parameter : <NR2>, Range: 0.400 ~ 0.750(AC RLC CP Mode)
Query Syntax
: [LOAD:] RLC:PFAC?
Response Parameter: <NR2>
Example
: RLC:PFAC 0.6
8.6.3.8 Inrush Current Subsystem
[LOAD:]
INRush
:CAP
:LS
:RL
:RS
:PHASe
[LOAD:] INRush:CAP
8-20
63802 Range: 2.77 ~ 9999.99, Unit: Ohm
Set the C
Set the L
S
Set the R
L
Set the R
S
Set the start phase
(AC RLC Inrush Current Mode)
(AC RLC Inrush Current Mode)
(AC RLC Inrush Current Mode)
(AC RLC Inrush Current Mode)
(AC RLC Inrush Current Mode)