Regenerative Charge/Discharge Tester 69225/69212 Series User's Manual
ABS(Vlocal-Vrmt)
WIR_LOS
ISET_MAX*0.03
Hardware protection
SYS_OCP
SYS_OTP
Hardware protection
SYS_FAN
Hardware protection
AUX_ERR
Hardware protection
Hardware protection
CAL_ERR
Software protection (Effective Program
Mode)
SW_OVP
Vlocal
ISET_MAX*0.03
Software protection (Effective Program
Mode)
SW_UVP
Vlocal
ISET_MAX*0.03
Note
1. DC Bus rated: 420V
2. VSET_MAX / ISET_MAX: Refer the section for specification.
3. Protections above are for single channel.
1.2.3
Output
Output connector is located at rear panel. Each output channel consists of upper
charge/discharge output and lower discharge output.
Manual operation needs to via Chroma 69200-x controller (optional equipment) to
proceed.
Remote operation needs to via Chroma 69200-x controller (optional equipment) to
coordinate with Ethernet card (optional equipment) and IPC via software to do remote
control.
Output protection detection (OUT_RVS, OUT_OVP, OUT_OCP, BAT_RVS, BAT_OVP,
WIR_LOS, etc.)
1.2.4
Input
Input connector is located at rear panel, each tester has one input connector which
provides 420V to do dual power transmission from A691101.
Input protection detection (BUS_OVP, BUS_UVP, etc.)
1.2.5
Measurement and Edit
System needs to via Chroma 69200-x controller (optional equipment) to coordinate with
software to perform outputting current/voltage/power... etc. measurements.
System needs to via Chroma 69200-x controller (optional equipment) to coordinate with
software to perform outputting Program, Waveform... etc. edits.
1-2
>
Vmax*5% +
≥
(OVP_SET + Vmax*5%) +
≤
(UVP_SET - Vmax*5%) -
Wire loss too large (or RMT wire off)
Hardware pulse by pulse overcurrent
protection.
692 inner temperature over high
692 inner fan blocked
Auxiliary power error is detected.
Calibration value error (uncalibrated
included)
Local voltage is higher than output
setting (user-defined)
Local voltage is lower than output
setting (user-defined)