Current Calibration; Aca Current Calibration - Chroma 19020 User Manual

Multi-channel sync hipot tester
Hide thumbs Also See for 19020:
Table of Contents

Advertisement

Multi-Channel Sync Hipot Tester 19020/19020-4/19021/19021-4/19022/19022-4
User's Manual
OSC 100V FULL (100V) Calibration Screen:
CALIBRATION
OSC 100V FULL (100V)
CHANNEL: 1
OUTPUT
100V
STANDARD
110V
DEF=100V
1.
Press START to output voltage and read the data from the high voltage meter.
2.
Press the 0
. - 9 numeric keys to input the reading from the high voltage meter.
3.
Press ENTER to confirm the input.
4.
Press STOP to stop the high voltage output.
5.
Press ▼ to go to the ACA 3mA OFFSET (0.12mA) calibration.

7.5 Current Calibration

The virtual load must be connected between the high potential terminal
CAUTION
and the ammeter input terminal to prevent a safety hazard.

7.5.1 ACA Current Calibration

(30uA and 300uA are for 19021, 19021-4 only)
Connect a 4.42pF capacitance load (60Hz equivalent impedance is 600M ) with a
20k
shunt resistance between the HV OUTPUT and the low potential terminal
(RTN/LOW) of the channel to be calibrated. Connect the handheld meter to the
Shunt resistor to measure the AC voltage.
7-10
READING
104V
REMOTE LOCK
NEXT CH
PREV CH
CORR
ERROR