SET UP
THERAPY SYSTEM SET UP (continued)
Setting System Volume
Press Volume button until
the desired system volume is
achieved. There are six settings:
Off, X-Low, Low, Med, High, and
X-High.
Each time the Volume button is
pressed the setting displayed will
emit three beep tones at that
level.
Ultrasound Coupling
This warning system works in conjunction with the Applicator
VOLUME
LED to alert the user should the Sound Head become uncoupled
BUTTON
from the patient. Press the US Coupling button until the desired
setting is displayed. There are four different alarm settings and an
Off setting.
Pause and Beep
Pauses Treatment Time and
emits an audible beep. When
the Applicator Sound Head is
re-coupled to the patient, the
Treatment Timer will automatically
restart.
Pause and No Beep
Pauses Treatment Timer. When
the Applicator Sound Head is
re-coupled to the patient, the
Treatment Timer will automatically
restart.
Beep
Emits an audible beep.
No Beep
No beep is emitted.
Off
Turns the Ultrasound
Coupling feature.
32
Intelect® Advanced Therapy System
US COUPLING
BUTTON