TROUBLESHOOTING
ERROR CODES (continued)
Code
Type
Number
Message
129
Warning
sEMG Data Card full.
130
Warning
Another treatment is running while attempting to set up and
perform a Laser Therapy treatment.
131
Warning
Treatment Room Door Lockout is breached.
132
Warning
Attempted to start a laser treatment but no Laser Applicator is
plugged in.
133
Warning
Laser Applicator became unplugged while performing a laser
treatment.
134
Warning
Entered incorrect laser PIN.
Probable Cause
110
Possible Remedies
sEMG Data Card faulty. Insert a known good sEMG Data Card. If problem continues, contact dealer
or Chattanooga Group for service.
A. Allow existing treatment to complete before starting Laser Therapy.
B. If no other treatment is running, reset Therapy System by turning main power switch
Off and On.
A. Make certain Treatment Room Door is completely closed.
B. Make certain the Lockout cable is connected to the system.
C. Replace Lockout to System cable with a known good cable.
D. Contact department responsible for installation of the Treatment Room Door Lockout
mechanism for maintenance or repair.
E. If problem continues, send Laser Module to Factory for service.
A. Connect desired Laser Applicator to the system.
B. If Applicator is connected, reset Therapy System by turning main power switch Off and On.
C. Connect a known good Laser Applicator.
D. If problem continues, send Laser Module to Factory for service.
A. Connect desired Laser Applicator to the system.
B. If Laser Applicator is connected, reset Therapy System by turning main power switch
Off and On.
C. Connect a known good Laser Applicator.
D. If problem continues, send Laser Module to Factory for service.
A. Enter correct Laser PIN number.
B. If problem continues, send Laser Module to Factory for service.