Chattanooga Intelect Operating Instructions

Chattanooga Intelect Operating Instructions

Advertisement

Quick Links

Ultrasound
1. Turn on unit (the main switch is on top of the main
screen on the left)
2. Select "Ultrasound" menu (this is the 2
console (first row)).
3. Press "Customize"
4. Set Frequency: 1 or 3 MHz
5. Set Mode: Continuous or Pulse mode (duty cycle
10%, 20%, or 50%)
6. Set display (W or W/cm
2
to display W/cm
7. Set time
8. Ignore Duty Cycle Frequency
9. Place the play button on the round dial.
10. Apply gel to the tissue and then apply the
transducer. To start the treatment, rotate the
Intensity knob to the desired output. The light on the
ultrasound head will turn green when the unit is
active.
11. If you want to stop treatment before treatment is
over, turn the dial down to 0, or press the stop
button below the round dial. You can also pause
treatment by pressing the pause button on the
round dial.
12. When treatment is completed, hold down the power
button until the unit turns off
myelectrotherapy.org
Chattanooga Intelect OPERATING INSTRUCTIONS
nd
key on the
2
). Ensure that it is selected
TENS
1. Turn on unit (the main switch is on top of the
main screen on the left)
2. Press Channel 1.
3. Select "Electrotherapy". This is the on the 1
of buttons on the console
4. Select "Pulsed Biphasic"
5. Select "symmetrical biphasic TENS"
6. Press customize:
a. Adjust the phase duration
b. Ignore frequency modulation
c. Ignore amplitude modulation
d. Adjust the frequency
e. Ignore beat frequency
f.
Adjust time
g. Leave to CC
h. Adjust the amplitude
i.
Once desired amplitude is reached, press
the play button
7. Press the back button
8. Press channel 2
9. Repeat steps 3-6
10. Press channel 1
11. Press play
12. When treatment is completed, hold down the
power button until the unit turns off
st
row

Advertisement

loading

Summary of Contents for Chattanooga Intelect

  • Page 1 Chattanooga Intelect OPERATING INSTRUCTIONS Ultrasound TENS 1. Turn on unit (the main switch is on top of the main 1. Turn on unit (the main switch is on top of the screen on the left) main screen on the left) 2. Select “Ultrasound” menu (this is the 2 key on the 2. Press Channel 1. console (first row)). 3. Select “Electrotherapy”. This is the on the 1 row 3. Press “Customize” of buttons on the console 4. Set Frequency: 1 or 3 MHz 4. Select “Pulsed Biphasic” 5. Set Mode: Continuous or Pulse mode (duty cycle 5. Select “symmetrical biphasic TENS” 10%, 20%, or 50%) 6. Press customize: 6. Set display (W or W/cm ). Ensure that it is selected a. Adjust the phase duration to display W/cm b. Ignore frequency modulation 7.
  • Page 2 Chattanooga Intelect OPERATING INSTRUCTIONS Interferential Current NMES 1. Turn on unit (the main switch is on top of 1. Turn on unit (the main switch is on top of the main screen on the main screen on the left) the left) 2. Select “Electrotherapy”. This is the on the 2. Press Channel 1. row of buttons on the console 3. Select “Electrotherapy” menu (this button is located on the 3. Select Interferential Current. first row of the console) 4. Select IFC 4 Pole 4. Select “Pulsed Biphasic” 5. Press customize: 5. Select VMS a. Adjust the carrier frequency (i.e. 6. Customize: 4000 Hz) a. Channel mode: single (using one channel), reciprocal b. Adjust the sweep to “on” (channel 1 goes on, turns off while channel 2 goes c.
  • Page 3 Chattanooga Intelect OPERATING INSTRUCTIONS Russian Current 1. Turn on unit (the main switch is on top of the main screen on the left) 2. Press Channel 1. 3. Select “Electrotherapy” menu (this button is located on the first row of the console) 4. Select “Russian Current” 5. Customize: a. Channel mode: single (using one channel), reciprocal (channel 1 goes on, turns off while channel 2 goes on. When channel 2 turns off, channel 1 goes on), or co-contraction (both channels go on at the same time). Select co-contraction. b. Select the cycle time including: i. the ramp time (how long it takes to reach maximal contraction, you can set this to 2 seconds ii. the on time which corresponds to the work time (how many continuous seconds it’s on for can set it to 10 seconds) iii. the off time, which corresponds to the rest time (how long it is off before it turns back on again, you can set it for 50 seconds) iv. Frequency. Select the appropriate frequency for strength or endurance v. Phase duration (which is also the pulse duration) set it to 300 microseconds vi. Time – total treatment time (if you want 10 contractions, and the work time is 10 seconds and the rest time is 50 seconds, then this treatment will take you 10- minutes) vii. CC/CV – leave to CC (constant current) 6.