TROUBLESHOOTING
ERROR CODES (continued)
Code
Type
Number
Message
118
Warning
Attempting to save additional User Protocols or Sequences after
system memory has reached the maximum allowed (200).
119
Warning
1. Attempted to read a treatment from Patient Data Card that is
120
Warning
2. Attempted to use a Non-Patient Data Card.
121
Warning
3. No Patient Data Card inserted into system port.
122
Warning
4. Unknown type of smart card inserted into system.
123
Warning
Patient Data Card is full.
124
Warning
Patient Treatment Data already saved.
125
Warning
Multimedia Card (MMC) not in system port.
126
Warning
No valid channels are available for attempted treatment.
127
Warning
1. No sEMG Channels are available for treatment.
128
Warning
2. No sEMG Module installed or detected by system.
Probable Cause
not a valid treatment for the system
Possible Remedies
Delete some User Protocols or Sequences. Refer to appropriate section of the Therapy System User
Manual for instructions.
A. Use a Patient Data Card with proper treatment data for the system.
B. Properly insert a Patient Data Card.
C. Insert a known good Patient Data Card.
D. If problem persists, insert a known good Patient Data Card. If problem continues, contact
dealer or Chattanooga Group for service.
Erase Patient Data Card. Refer to Therapy System User Manual for instructions.
A. Cannot save same data again on Patient Data Card.
B. Use a new Patient Data Card to resave data.
C. Erase Patient Data Card and resave treatment data.
A. Properly insert the MMC card into the system port.
B. Insert a known good MMC Card. If problem continues, contact dealer or Chattanooga Group
for service.
A. Complete existing treatment before attempting to start another.
B. Reset Therapy System by turning main power switch Off and On.
A. Wait until current treatment is complete.
B. Reset Therapy System by turning main power switch Off and On.
C. Make certain sEMG Module is properly installed. Refer to sEMG Module User Manual for
installation instructions.
D. Replace sEMG Module with known good sEMG Module.
E. If problem continues, contact dealer or Chattanooga Group for service.
109
Intelect® Advanced Therapy System