INSTALLATION/REMOVAL
INSTALLATION CHANNEL 3/4 ELECTROTHERAPY MODULE (continued)
Turn Therapy System On
Turn the System On using the On/Off
Switch. The System will automatically
recognize the added Module and display a
configuration change message.
Read and carefully follow the instructions
on the screen.
Verify that the Module installed is the Module
displayed in the message BEFORE pressing
the START button. If it is not, DO NOT press
the START button. Turn the System OFF
and back ON. If the problem continues, call
the selling dealer or Chattanooga Group
Technical Support immediately.
DO NOT USE THE SYSTEM until all necessary
repairs are made by a Technician certified
by Chattanooga Group. If use is attempted
before repairs are made, the System may
operate unpredictably and has the potential
of causing injury to the patient or damage to
the System's internal components.
102
Intelect® Advanced Therapy System