Usage Information
Setup and Use of Keysight 855xxA/B Series CalPods
2. To access the Windows desktop when the PNA application is active, select File >
Minimize Application.
3. Right-click on Start. and select Explore.
4. Navigate to the files on the USB drive, and double-click setup.exe to run the
program.
5. In the dialog window titled "CalPod Thermal Characterization Data Install
Utility," click "I accept license terms..." and wait for additional text to be
displayed. This loads CalPod Temperature Characterization data in the PNA.
6. When "PNA restore archive complete" is displayed at the bottom of the
window, close the window.
7. Remove the USB drive from the PNA.
8. Repeat
each CalPod to be used with the PNA.
Figure 1-1
USB Drive Matches Serial Number on CalPod

Step 7. Connect the LAN Cable

A LAN cable is required but is not included. Either a standard or cross-over LAN
cable can be used. (A cross-over LAN cable is used to connect computing devices
together directly where they would normally be connected via a network switch,
hub, or router.)
— The following graphic shows the recommended configuration for the LAN cable.
Information on an alternative configuration can be found at
85563-90005 Operation & Service Guide
"Step 6. Load Temperature Characterization Data" on page 1-11
for
"Alternative
1-13