Solvent Information - Agilent Technologies 1290 Infinity LC System System Manual And Quick Reference

Table of Contents

Advertisement

Solvent Information

Flow Cell
To protect optimal functionality of your flow-cell:
• Avoid the use of alkaline solutions (pH > 9.5) which can attack quartz and
Use of Solvents
Observe the following recommendations on the use of solvents.
• Brown glass ware can avoid growth of algae.
• Avoid the use of the following steel-corrosive solvents:
Agilent 1290 Infinity LC System Manual and Quick Reference
thus impair the optical properties of the flow cell.
Solutions of alkali halides and their respective acids (for example,
lithium iodide, potassium chloride, and so on),
High concentrations of inorganic acids like sulfuric acid and nitric acid,
especially at higher temperatures (if your chromatography method
allows, replace by phosphoric acid or phosphate buffer which are less
corrosive against stainless steel),
Halogenated solvents or mixtures which form radicals and/or acids, for
example:
→ 2COCl
2CHCl
+ O
3
2
This reaction, in which stainless steel probably acts as a catalyst, occurs
quickly with dried chloroform if the drying process removes the
stabilizing alcohol,
Chromatographic grade ethers, which can contain peroxides (for
example, THF, dioxane, di-isopropylether) such ethers should be filtered
through dry aluminium oxide which adsorbs the peroxides,
Solvents containing strong complexing agents (e.g. EDTA),
Mixtures of carbon tetrachloride with 2-propanol or THF.
+ 2HCl
2