2 In the Purge section, set the following parameters:
3 Right-click on the pump section, and select Purge On from the context menu.
Do not confuse Purge On with the next item Prime On.
N O T E
The purge valve will now switch the flow path to the purge position, and will
simultaneously purge both channels to waste at 5 ml/min each for 6 minutes
duration. After the set duration, the purge flow will turn off, the valve will
switch the flow path back to the system, and the flow rate and composition
currently set in the method will resume. In this example, the method flow is
still set to zero. The purge settings in the Control page from steps 1 and 2
remain configured so when the purge process is required again it can be
started at step 3.
When the pump has been initially primed and purged with propan-2-ol, the
solvents can be exchanged for the mobile phase solvents such as water and
methanol. The purge procedure is repeated whenever a solvent is exchanged.
Immediately following the purge, the solvent in the pump has not been
degassed and so the system should be operated for at least 10 min so that the
solvent is degassed.
Agilent 1290 Infinity LC System Manual and Quick Reference
Duration: 6 min
Flow: 10 ml/min
Composition B: 50 %
Composition A will automatically assume 50 %. Leave the On/Off button set to
Off. Click OK to exit.
System Setup and Installation
Installing the Module
4
87