3
Optimization of the Agilent 1290 Infinity LC System
How to Configure the Optimum Delay Volume
applications there is no need for an additional Jet Weaver mixer. But it is still
possible to install an optional 380 µL Jet Weaver mixer for baseline critical
applications, such as TFA applications. The optional Jet Weaver mixer has a
different housing, which is adapted to the design of the 1290 Infinity
Quaternary Pump.
The procedure for installing a Jet Weaver in the 1290 Infinity Quaternary
Pump is illustrated in
"Installing the V380 Jet Weaver in the 1290 Infinity
Quaternary Pump"
on page 49.
Agilent 1290 Infinity LC System Manual and Quick Reference