Agilent Technologies 1290 Infinity LC System System Manual And Quick Reference page 123

Table of Contents

Advertisement

Binary Pump (BinPump) tab
Figure 39
• Flow sets the flow rate up to 5 ml/min. For the example separation
• Solvents defines the mobile phases that are available and the percentage
Agilent 1290 Infinity LC System Manual and Quick Reference
Setup Method screen – Binary Pump tab
0.4 ml/min is used. If the back pressure briefly reaches the maximum
pressure setting the flow will be reduced for a few seconds to lower the
pressure but if the pressure continues to be limited in this way an error
condition will be created and the flow will be stopped.
proportions pumped on the two channels, A and B. On each channel a
drop-down box allows the selection of a solvent from a list so that the pump
control uses the optimum compressibility settings. This will optimize the
flow characteristics as described in
Volume"
on page 42. A second text box allows a description of the mobile
phase to be entered. If the solvent selection valve is installed in the pump
then each channel has two solvent options and the correct option for the
method is selected with the radio button to the left of the solvent
description. The pump will form binary mixtures of the selected A and B
Setting Up a Method using Edit Entire Method
"How to Configure the Optimum Delay
6
Appendix
123