Agilent Technologies 1290 Infinity LC System System Manual And Quick Reference page 128

Table of Contents

Advertisement

6
Appendix
Setting Up a Method using Edit Entire Method
Diode-Array Detector (DAD) tab
Figure 41
• Signals: Up to eight separate signals (chromatograms) can be recorded. To
• Peakwidth sets the data collection rate and signal filtering.
• Stop Time / Post Time are set to No Limit / Off and these values are usually
128
Setup Method screen – Diode-Array Detector tab
mark a signal for collection the Use Signal box is checked for that signal, the
wavelength and bandwidth are defined, and if a reference signal is required
that box is checked and defined.
Wavelength sets the central wavelength (nm) of the signal,
Bandwidth sets the width (nm) of the signal,
Reference Wavelength sets the central wavelength (nm) of the reference
band which is subtracted from the analytical signal,
Reference Bandwidth sets the width (nm) of the reference band.
taken care of in the Pump tab. However, the detector module Stop Time can
sometimes differ from the pump Stop Time when it is required that data
Agilent 1290 Infinity LC System Manual and Quick Reference